pNeuton mini NEO Ventilator with CPAP for invasive and non-invasive patient support anywhere
The pNeuton mini NEO ventilator is a safe and secure way for neonatology to transport and perform MRI procedures specifically designed for neonatal ventilation.
MRI Compatible
The pNeuton mini NEO is MRI compatible allowing for image production with no artifact. The
ventilator can be placed right next to the MRI magnet with no Gauss line restriction, plus a remote alarm can be added to enhance patient safety.
Specs
Description
-
Pneumatically powered for use on neonates and infants
-
Patient range: 400 gram to 20 kilogram
-
Modes: CMV, IMV + CPAP or CPAP only – continuous flow pressure limited
ventilation
-
miniFlow Patient Interface supports nasal prong/mask application
-
Pressures displayed on manometer
-
MRI conditional: static magnetic field of 3 T or less, maximum spatial gradient
magnetic field of 720-gauss/cm or less, no gauss line restriction
-
Gas consumption: flow setting + 3 L/min oxygen
-
Weight: 9 lbs (4 kg)
-
Dimensions: 6.0″ h x 8.7″ w x 7.8″ d
(15.2 cm x 22.1 cm x 19.8 cm)
-
Input gas requirement (oxygen and medical air): 55 psi ± 15 psi (3.8 bar ± 1 bar)
each gas
-
Meets International Standards for transport ventilators:
-
ASTM F1100-90 – Ventilators Intended for Use in Critical Care
-
ISO 10651-3:1997 – Lung Ventilators For Medical Use – Particular requirements for emergency and transport ventilators
-
Airworthiness: RTCA DO-160G – Environmental Conditions and Test Procedures for Airborne Equipment, as applicable
Control Settings
Inspiratory time 0.25 to 0.8 seconds
Expiratory time 0.25 to 6 seconds
Continuous Flow 6, 8, or 10 L/min
Peak pressure 15 to 40 cm H2O (mbar)
PEEP / CPAP 0 to 20 cm H2O (mbar)
Oxygen 21 to 100% ± 3%, requires oxygen and medical air source
Audible and Visual Alarms
-
All pneumatic alarm system (no batteries – requires minimum 30 psi / 2 bar gas supply pressure)
-
Remote alarm output connection
-
Patient circuit disconnection
-
Automatic reset when alarm condition resolves
-
10-second response, 25-second silence/reset button
-
Pressure: less than 3 cm H2O
-
High pressure – independently adjustable from Peak Pressure
-
Low gas source pressure
-
If either source gas drops below 40 psi (2.8 bar)
-
Continues operation as long as oxygen is available
Please contact us for details.